1) The MS-200 OES is mainly used for the analysis of multiple metal matrix composites, including that of Nitrogen with super low content with the help of maximum 80 optical channels.
2) Digital-controlled light source with customizable excitation parameters is applied to different metals. The resultant analysis is more stable and can meet the analytical demand for trace elements and elements with super-high content.
3) Brand-new standardized module design; highly-integrated circuit; independent charging system;
4) All exit slits are on one belt with easy addition if needed.
5) Parts with high quality: core parts from the world well-known manufacturers.
6) Self-protection for optical system is in place through a built-in thermostat system.
7) Multi-elemental analysis is available for Fe, Al, Cu, Ni, Pb, Zn, Mg, Co, Sn, Ti.
8) Oil back-streaming prevention is achieved by creating a vacuum environment in the optical chamber, making no vaporized oil come through. A computer-managed vacuum pump enables its working time less than 5% of the total MS-170 working time, largely extending the pump life.
9) Highly-precise thermostatic system can keep the range of temperature fluctuation between -0.2℃ and +0.2℃.
10) Automatic plotting technology is applied to optical calibration to position smartly the entrance slit.
11) 10-20s of analytical process
12) User-friendly operational software with multiple language versions under WINDOWS operating system.
13) The multipurpose operational software can meet your various needs by displaying elemental data in a flexible way and supporting various printing formats.
14) SSA can make happen the analysis of acid-soluble and acid-insoluble substances.
Availability: | |
---|---|
MS-200
Mikrosize
90273000.00
Product Description
optical system | structure | Paschen-Runge |
focal length | 750mm | |
low-noise holographic original blaze grating | 3600 ruled lines/mm | |
wavelength range | 120-800nm | |
dispersive power | first order 0.37nm/mm | |
vacuum optical chamber with auto thermostatic system | 30±0.2℃ vacuum range: 1.2-2.5 Pa | |
light source system | light source | multifrequency digital controlled light source SSA |
discharge current | 10-400A | |
excitation frequency | 100-1000Hz | |
discharge duration | 10-10000μs | |
spark stand | excitation chamber with minimum argon use | |
easily-changeable foundation(lid) | ||
mobile hold-down pin | ||
data collection and control system | dynamic integration technology | |
highly-sensitive PMT detector | ||
high-speed 16-bit A/D conversion | ||
real-time control on temperature and vacuum state | ||
SSA | ||
ethernet, serial port, USB | ||
others | matrices | Fe, Cu, Al, Ni, Zn, Mg, Pb, Co, Sn, Ti, etc. |
application | trace, micro and macro elements, N element pure metals | |
channel configuration | 6 matrices and 80 channels at most | |
dimensions(mm) | 1388*910*1210 | |
environmental requirement | T 10℃-35℃ H 20%-80% | |
weight | 400Kg (net weight), 470Kg (gross weight) | |
V/F | AC220V±10%/50Hz | |
power | Max 1500VA, standby 300VA | |
argon | purity≥99.994% press≥0.3MPa |