Product Feature:
● The sulfur dioxide salt spray tester is suitable for corrosion testing of metal material protection, parts, electronic devices and industrial products, and can quickly and conveniently reproduce the corrosion process of metal surfaces after spraying and without treatment.
● It can meet the test requirements of GB9789-1988 "Metallic and other non-metallic organic coatings - Sulfur dioxide corrosion test under typical condensation conditions" and related standards for various corrosion gas tests.
● The standards this equipment complies with are: GB/T2423.17-2008/1EC 60068-2-11-1981 ASTM.B117-2009 JIS H8502 GB/T10125-2012/1SO 9227-2006 GB-T5170.8-2008 GB/T5170.11-2008 GB/T10587-2006 GB/T2423.33-2005 DIN 50188-1997.
Product Application:
● This test equipment simulates corrosion tests in artificial environments, suitable for determining the adaptability of electrical and electronic products and their materials for use under chemical corrosion conditions and for comparing the quality of metal materials with or without corrosion protection.
● It is suitable for corrosion testing for the rapid analysis of discontinuities, pores and damage in organic and inorganic coatings.
Availability: | |
---|---|
Quantity: | |
Product Description
Technical Specification:
Name | Cyclic Corrosion Test Chamber
| |||
Model | ASR-FH-90D | ASR-FH-120D | ASR-FH-160D | ASR-FH-200D |
Internal Dim. (mm) ( W * D * H)
| 900x600x500 | 1200x1000x500 | 1600x1000x500 | 2000x1000x500 |
External Dim. (mm) ( W * D * H)
| 2450x1553x 1430 | 2830x1560x 1830 | 3230x1560x1830 | 3630x1560x 1930 |
Temperature Range | 10℃-70℃ | |||
Humidity Range | 30%~90% | |||
Temperature and Humidity Analysis Precision/Uniformity | 0.1℃;0.1R.H./+1.0℃;+3.0%R.H. | |||
Temperature and Humidity Analysis Accuracy/Fluctuation | +0.1℃;+2.0R.H./+0.5℃℃;+2.0%R.H. | |||
Laboratory Temperature | Saltwater test method(NSSA CSS)35℃+1℃ Corrosion Resistance Test Method(CASS)50℃±0.1℃ | |||
Pressure Tank Temperature | Saltwater test method(NSSACSS)47℃+1℃ Corrosion Resistance Test Method(CASS)63℃℃±0.1℃ | |||
Saltwater Temperature | 35℃+1℃ or 50℃+1℃ | |||
Saltwater Concentration | The concentration of sodium chloride solution is 5% or 0.26 grams of copper chloride is added per liter of sodium chloride solution with a concentration of 5%. | |||
Compressed Air Pressure | 1.00+0.01kgf/cm | |||
Spray Volume | 10~2.0ml/80cm²/h(The standard requires 16 hours of testing for an average quantity) | |||
PH Value | 6.5~7.23.0~3.2 | |||
Spray Method | Using Bernoulli's principle to absorb brine and then atomize it. | |||
Power Supply | 380V 6KW 50HZ MAX 30A | |||
Salt Spray Test | Salt Spray Test | (1) Laboratory: 10℃~60℃ (non-linear) (2)Saturated air barrel:47℃ ±1℃ | ||
Wetting Test | (1) Temperature: 49℃ ±2℃ (2) Humidity: 30%~95% ±3% (3)Saturated air barrel:47℃ ±1℃ | |||
Drying Test | (1)Temperature: 10℃~65℃ ±0.5℃ (2) Humidity:More than 93% | |||
Circuit Control System | 1.The main controller is equipped with a 7-inch LCD human-machine interface, and the main screen can be planned out with the test procedure. 2.Fixed conditions can set time: 0 - 999 H 59 MIN。 3.Program mode can set salt spray and humidity test at the same time 4.Equipped with power-off memory device, the remaining program can be continued when power is restored. 5.The temperature rise and fall can be controlled at a rate, and has the function of self-correcting the temperature and humidity reference points. 6. The temperature and humidity sensor uses a DIN PT-100Ω (platinum sensing).
| |||
Air Supply System | The air pressure is 1kg/cm2 and is adjusted in two stages. The first is the imported pressure regulating filter with water filtering function, which is roughly adjusted to 2-2.5kg/cm2. The second is the imported precision pressure regulating valve, which is adjusted to the pressure gauge 1kg/cm2. | |||
Inlet Drainage System | The system is equipped with an automatic water replenishment feature that adds water when the level is too low, and manual water addition is also available. | |||
Humidification System | 1. Adopts electronic bit P.I.D micro-motion humidification system. 2. With overheating and overflow dual protection device 3. The humidification and dehumidification systems are completely independent. 4. Water level control using high molecular float ball valves. 5.Automatic Humidifier Water Refill Device. (Machine needs to be connected to a water source.)
| |||
Heating System | 1.It adopts external hot air heating method, which heats up quickly, reduces standby time, and has a low component damage rate. 2.Use S.S.R to control heating power, accurate temperature and low power consumption.
| |||
Dehumidification System | 1.The refrigeration evaporator condensation dehumidification method is adopted to ensure stable humidity. 2.Bubble accelerated humidifier, humidity stabilization rate is fast. 3.Adopts a high-speed return air suction system.
| |||
Safety Protection Device | 1.Equipped with leakage circuit breaker without fuse protection. 2.Low water level protection switch. 3.Low-salt solution protection switch 4. Dual overheat protection device, preventing mechanical or electronic actions from running out of control. 5. Fault point Chinese indication screen.
| |||
Using Environment | Performance guaranteed environmental temperature range:+5℃~+30℃(Except for the minimum attainable temperature, maximum thermal load capacity and cooling time) |
Packing List:
Name | Quantity |
Shelf | 6 pcs |
Standard Measuring Cylinder | 2 pcs |
Glass Filter | 2 pcs |
Collector 80cm2 | 2 pcs |
Glass Nozzle | 2 pcs |
Sodium Chloride (NACL) 500 g | 2 bottles |
Instruction Manual | 1 copy |
Product Certificate | 1 copy |
Warranty Card | 1 copy |